Connecting The Apci Probe To The Ms Detector - Thermo Scientific LCQ Fleet Getting Started Manual

Ltq series
Table of Contents

Advertisement

2
Connecting Probes

Connecting the APCI Probe to the MS Detector

Connecting the APCI Probe to the MS Detector
12
LTQ Series Getting Connected Guide
To connect liquid lines to the atmospheric chemical ionization (APCI) probe
1. Install the Ion Max source housing and APCI probe onto the LTQ Series mass
spectrometer as described in the Ion Max and Ion Max-S API Source Hardware Manual.
2. Install liquid lines between the divert/inject valve, the LC system, the syringe pump, and
the sample inlet fitting on the APCI probe, as is appropriate for your application. For
more information, see
3. Connect the 1-in. ID Tygon™ tubing (P/N 00301-22922) to the source housing drain.
4. Insert the other end of the tubing into a waste container, and vent the waste container to
a fume exhaust system.
CAUTION Prevent solvent waste from backing up into the API source and mass
spectrometer. Always ensure that the PVC drain tubing is above the level of liquid in
the waste container.
Do not vent the PVC drain tube connected to the waste container to the same fume
exhaust system to which you have connected the forepump. Connecting the API
source drain tube and the forepump exhaust to the same fume exhaust system is likely
to contaminate the analyzer optics.
Equip your laboratory with at least two independent fume exhaust systems. Route the
(blue) exhaust tubing from the forepump to a dedicated fume exhaust system. Route
the PVC drain tube from the API source to the waste container. Vent the waste
container to a dedicated fume exhaust system that is independent from that used for
exhausting the forepump.
Note If you need to install or replace the APCI sample tube, refer to the hardware
manual for your Thermo Scientific mass spectrometer.
Chapter 6, "Connecting the Inlet Plumbing."