Sensitivity Adjustment] Screen - A&D BA-T Series Instruction Manual

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

10. [Sensitivity adjustment] screen

Display the [Sensitivity adjustment] screen as follows:
Display settings: MENU key
1
Int. sensitivity adjustment button
2
Ext. sensitivity adjustment button
Name
1
Int. sensitivity adjustment button
2
Ext. sensitivity adjustment button
Automatic sensitivity adjustment
3
button
Sensitivity adjustment setting
4
button
Since the balance's resolution is high, weighing values may change due to gravity and daily
environmental changes. It is necessary to perform sensitivity adjustment with the weight in order to keep
the weighing values from changing even if gravity or the environment changes.
It is recommended that you perform sensitivity adjustment if the balance is installed for the first time or
relocated, or when the weighing values change significantly in daily check, etc.
Sensitivity adjustment means to adjust the weighing value of the balance using the reference weight or
internal weight.
Caution on sensitivity adjustment
Do not allow vibration, drafts or temperature changes to affect the balance especially during sensitivity
adjustment.
The GLP/GMP (etc.) compliant report can be output in sensitivity adjustment. A computer or optional
printer is required for GLP output. A timestamp (clock and calendar) is available for the GLP output
using the clock function of the balance.
You can set the clock function from the [Date/time setting] screen.
Display settings: MENU key
[Sensitivity adjustment] screen
Displays the [Internal sensitivity adjustment] screen to execute the internal
sensitivity adjustment.
After adjustment, the [Sensitivity adjustment result] screen for the internal
sensitivity adjustment is displayed.
Displays the [External sensitivity adjustment] screen to start the external
sensitivity adjustment.
After adjustment, the [Sensitivity adjustment result] screen for the external
sensitivity adjustment is displayed.
Displays the [Automatic sensitivity adjustment] settings screen.
The automatic sensitivity adjustment is executed.
Displays the [Sensitivity adjustment setting] screen. You can configure the
settings for the CAL button, standard value for external weight values and
internal weight value correction for the weighing screen.
[Date/time setting] screen
Sensitivity adjustment button
Description
System settings button
62
3
Automatic sensitivity adjustment button
4
Sensitivity adjustment setting button
Date/time button