A&D BA-T Series Instruction Manual page 86

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Repeatability check
Display settings: MENU key
1
Measurement method
2
Measurement count
3
Tolerance (standard deviation)
Name
Measurement
1
method
Measurement count
2
Tolerance (standard
3
deviation)
Measure button
4
Back button
5
Settings in the red box are default values (factory settings).
For the internal weight, the repeatability check is automatically performed.
For an external weight, follow the instruction.
When the measurement is complete, [Repeatability measurement result] is automatically displayed.
This is common to the [Quick performance check: Repeatability measurement] screen described in
"6. Quick performance check: [Repeatability Measurement]
Repeatability measurement with the internal weight
Display settings: MENU key
1
Name
Back button
1
Repeatability measurement result
2
display
PRINT button
3
Exit button
4
Calibration test/check button
Repeatability check button
Setting value (setting range)
Automatic (internal weight) ,
Manual (external weight)
Calibration test/check button
Repeatability check button
with [Automatic (internal weight) selected for the measurement method, select
Measure button
Returns to the previous screen.
Displays the repeatability measurement result.
Outputs the result to the device connected to the balance.
Exits the screen that shows the repeatability measurement result.
[Repeatability measurement] screen
Select the measurement method.
-
Displays the measurement count.
-
Displays the tolerance (standard deviation).
-
Executes the measurement.
-
Returns to the previous screen.
Screen".
[Repeatability measurement] screen. In this screen,
[Repeatability measurement] screen with the internal weight
2
3
4
86
4
Measure button
Description
Description

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte