4. Weighing pan check
Make sure that there is nothing on the weighing pan.
1
Skip button
2
Back button
Name
Skip button
1
Back button
2
Confirm button
3
5 Accuracy check
Use your test weight to make sure that the deviation falls within the tolerance.
1
2
3
Name
Test weight value input
1
field
Tolerance input field
2
Back button
3
Skip button
4
Accuracy check
5
Start button
Test weight value input
6
field
Accuracy check result
7
display
Judgment
8
Next button
9
Skips this check item and proceeds to the next item.
Returns to the previous screen.
Considers this check item as OK and proceeds to the next item.
4
5
Input the value for your test weight.
Input the tolerance.
Returns to the previous screen.
Skips this check item and proceeds to the next item.
Displays the [Accuracy check mode] screen.
Input the value for your test weight.
This is common to the one in the [Accuracy check] screen.
Displays the accuracy check result.
Displays the judgment.
Proceeds to the next check item.
Description
Description
72
3
Confirm button
6
7
8
9

Advertisement