A&D BA-T Series Instruction Manual page 42

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Automatic (internal weight)
For this setting, the minimum weight is automatically measured with the internal weight in the
[Repeatability measurement mode] screen. When the measurement is complete, the [Minimum weight
measurement result] screen described in
1
Back button
Name
Back button
1
Manual (external weight)
For this setting, measurement is performed with your weight to compute the minimum weight.
The procedure is displayed in the message field.
Weighing example for BA-6DTE
Step 1.
Press the RE-ZERO button
Step 2.
Place a weight on the weighing pan.
Step 3.
Remove the weight when the stabilization
indicator
Step 4.
Repeat steps 1, 2 and 3 ten times.
Step 5.
When the measurement is complete, the
[Minimum weight measurement result] screen
described in
measurement
Returns to the previous screen.
is displayed.
"5.5.4. Minimum weight
result" is displayed.
"5.5.4. Minimum weight measurement
Description
Weighing pan
.
RE-ZERO button
Place the weight
Remove
the weight
10 times complete
[Minimum weight measurement result] screen
42
result" is displayed.