Minimum weight measurement result
The [Minimum weight measurement result] screen is displayed when the
measurement
is complete.
1
Measurement method
2
Measurement tolerance
3
Repeatability (standard deviation)
4
Minimum weight
Name
Measurement method
1
Measurement tolerance Displays the minimum weight measurement tolerance.
2
Repeatability (standard
3
deviation)
Minimum weight
4
PRINT button
5
Register button
6
*
If the repeatability (standard deviation) is 0.41d or less, the minimum weight is 2000 times the 0.41d in
compliance with USP.
"d" is a unit of readability.
Displays the method used for the minimum weight measurement.
Displays the standard deviation for this measurement.
Displays the minimum weight.
Outputs the minimum weight measurement result to the device connected to
the balance.
Registers the minimum weight.
Description
43
5.5.3. minimum weight
5
PRINT button
6
REGISTER button