A&D standard format
This is the standard format for sending data to peripheral devices.
Consists of 16 characters (not including the terminator).
The condition of the data is indicated with a 2-character header.
The data is padded with polarity and zeros (filling the data's higher order's surplus part with zeros).
If the data is zero, the polarity is positive.
The unit consists of three characters.
,
S
T
+
Header
S
T When stable
U
S When unstable
T Counting mode when stable
Q
O
L When overloaded
In the external key print mode of the AD-8127 multi-functional
compact printer, a received A&D standard format is printed as shown
on the right.
If the gross/tare weight is added in the [Data to be added] screen of the function table, the second
header corresponding to the output data is appended to the header.
,
S
T
N
2nd header
N
G
T
P
Display settings: MENU key
.
0
1
2
3
Data
,
+
0
1
Net weight
Gross weight
Tare weight
T Preset tare weight
System settings button
Data output button
[Data to be added] screen
4
5
6
7
: Carriage return,
C
R
L
: Line feed,
F
: Space,
.
2
3
4
5
Data
Data to be added button
142
g
C
L
R
F
Terminator
Unit
ASCII 0Dh
ASCII 0Ah
ASCII 20h
6
7
g
Unit
: Carriage return,
C
R
L
: Line feed,
F
: Space,
Communication button
C
L
R
F
Terminator
ASCII 0Dh
ASCII 0Ah
ASCII 20h

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte