A&D BA-T Series Instruction Manual page 94

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Standard value setting: Sensitivity test
Display settings: MENU key
1
Number of loading points
for measurement
2
Measuring load 1
3
Measuring load 2
4
Back button
Name
Number of loading
points for
1
measurement
Measuring load 1
2
Measuring load 2
3
Back button
4
Settings in the red box are default values (factory settings).
You can change the sensitivity test settings.
The settings in this screen are common to the sensitivity test settings screen for the periodic check.
Standard value setting: Eccentricity
Display settings: MENU key
1
Tolerance (maximum deviation
from the center value)
2
Back button
Name
Tolerance (maximum
deviation from the center
1
value)
Back button
2
You can change the eccentricity settings.
The settings in this screen are common to the eccentricity settings screen for the periodic check.
Calibration test/check button
Standard value setting button
[Standard value setting: Sensitivity test] screen
Setting value
(setting range)
2 point , 1 points
-
-
-
Calibration test/check button
Standard value setting button
[Standard value setting: Eccentricity] screen
Set the tolerance (maximum deviation from the center value).
Returns to the previous screen.
Sensitivity test button
Set the number of loading points for measurement.
Set the nominal value and tolerance.
Set the nominal value and tolerance.
Returns to the previous screen.
Eccentricity button
Description
94
Settings button
Description
Settings button