A&D BA-T Series Instruction Manual page 64

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

External sensitivity adjustment
Display settings 1: MENU key
Ext. sensitivity adjustment button
Display settings 2: HOME key
Ext. sensitivity adjustment button
1
2
3
Name
Instruction display
1
External weight
2
value input field
Back button
3
Confirm button
4
Date
5
Time
6
Weight used
7
Adjustment weight
8
value
PRINT button
9
Settings in the red box are default values (factory settings).
This function performs sensitivity adjustment of the balance using your
external weight.
Be sure to warm up the balance with nothing on the weighing pan for at least an hour, or at least four
hours for BA-6TE/6DTE, with the AC adapter connected to the power supply.
Do not apply vibration and the like to the balance during sensitivity adjustment.
When the adjustment is complete, the [Sensitivity adjustment result] screen for the external sensitivity
adjustment is automatically displayed.
Caution
The accuracy of the weight used in sensitivity adjustment affects the accuracy of the balance after
sensitivity adjustment.
Sensitivity adjustment button
CAL button
4
Displays the instruction for external sensitivity adjustment.
Input the external weight value. This is common to the external weight value
input field in the [Sensitivity adjustment result] screen for external sensitivity
adjustment.
Input range
BA-6TE/BA-6DTE
BA-225TE/BA-225DTE
BA-125DTE
Returns to the previous screen.
Confirms the current input setting and proceeds to the next instruction.
Displays the date when the external sensitivity adjustment was complete.
Displays the time when the external sensitivity adjustment was complete.
Displays the type of the weight used.
Displays the weight value used.
Outputs the sensitivity adjustment result.
The result is automatically output if [GLP output] in [Data output] is set to ON.
[External sensitivity adjustment] screen
[External sensitivity adjustment] screen
Description
:
0.9 g ~
:
9.9 g ~ 200 g ~ 200.1 g
:
9.9 g ~ 100 g ~ 100.1 g
64
5 g ~
5.1 g
5
6
7
8
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte

Table of Contents