10. Eccentricity
The eccentricity is measured.
Follow the instruction for measurement.
The setting in the [Sensitivity test] screen is common to the standard value setting for the daily/periodic
check settings.
When the measurement is complete, the [Eccentricity measurement result] screen is automatically
displayed.
The accurate positions for measurement positions 2, 3,
diagonal line or diameter of the weighing pan.
It is advisable to use a weight approx. 1/3 as heavy as the capacity of the balance.
1
2
3
Name
Settings display screen
1
Tolerance (maximum
deviation from the center
2
value)
Back button
3
Eccentricity measurement
4
result display
Settings button
5
Measure button
6
Skip button
7
Weighing value
8
Next button
9
Weight position image
10
Displays the set tolerance (maximum deviation from the center value).
Set the tolerance for eccentricity.
Cancels and returns to the previous screen.
Cancels the eccentricity measurement and returns to the previous screen.
Displays the [Standard value setting] screen.
Configure the set tolerance (maximum deviation from the center value).
Executes the measurement.
Skips this check item and proceeds to the next item.
Displays the measurement positions and weighing values.
Proceeds to the next item.
Displays the positions where the weight is placed during measurement.
82
4
and
5
are at the 1/4 position of the
4
5
6
7
8
10
Description
9