Caution
If something is on the weighing pan, the balance itself will judge that it is in use and will not perform
automatic sensitivity adjustment.
The criteria for performing automatic sensitivity adjustment are as follows.
BA-6TE / BA-6DTE
BA-225DTE / BA-125DTE / BA-225TE
To maintain the correct sensitivity adjustment of the balance, do not place anything on the weighing pan
while not in use.
Sensitivity adjustment setting
Display settings: MENU key
[Sensitivity adjustment setting] screen
1
External weight value input
2
Enter the correct internal weight value button
External weight value input
1
Enter the correct internal weight value
2
button
Settings in the red box are default values (factory settings).
*1
Input range
BA-6TE/BA-6DTE
BA-225TE/BA-225DTE
BA-125DTE
Sensitivity adjustment button
Name
: 0.9 g ~
: 9.9 g ~ 200 g ~ 200.1 g
: 9.9 g ~ 100 g ~ 100.1 g
Less than 0.1 g
Setting value
(setting range)
Input the external weight value. This is
common to the "Adjustment weight value"
Input range
*1
input item in the [External sensitivity
adjustment] screen.
Displays the [Internal sensitivity
adjustment] screen.
5 g ~
5.1 g
67
Settings button
Description