A&D BA-T Series Instruction Manual page 32

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Smart range function
For BA-6DTE / BA-225DTE / BA-125DTE, there are two types of ranges: the standard range and
precision range (high resolution).
Smart range function
The range can switch automatically between the standard range and precision range (high resolution),
depending on the value displayed.
Pressing the RE-ZERO button
precision range, regardless of the tare value. The range can be fixed to the standard range by using
[Minimum display digit] in the [Application] settings screen.
Display settings: HOME key
Operation example
Example weighing in g for BA-6DTE
Step 1.
Start weighing in the precision range.
Press the RE-ZERO button
the display to zero and enable the
precision range.
Step 2.
Place a container on the weighing pan.
When the displayed value exceeds the
precision range, the balance automatically
switches to the standard range.
Step 3.
Press the RE-ZERO button
the display to zero and enable the
precision range.
Step 4.
Place a sample in the container.
If the weighing value does not exceed
the precision range, the sample can be
weighed in the precision range.
Step 5.
Remove the sample and container from
the weighing pan.
Precision range and standard range
Model
Unit
g
gram
BA-6DTE
mg milligram 0.000 mg
g
gram
BA-225DTE
mg milligram 0.00 mg
g
gram
BA-125DTE
mg milligram 0.00 mg
to set the displayed value to zero allows for weighing in the
APP button
Minimum display digit button
to set
to set
Precision range after pressing the
RE-ZERO button
0.000000 g
-
0.00000 g
-
0.00000 g
-
Select from [No blank], [1 digit blank], [2 digit blank].
Weighing pan
RE-ZERO button
Container (tare)
RE-ZERO button
Sample
Nothing on the pan
-
2.100009 g
2100.009 mg
-
51.00009 g
51000.09 mg
-
51.00009 g
51000.09 mg
32
[Application] settings screen
Zero display
Precision range
Standard range
Zero display
Precision range
Precision range
Standard range
2.10001 g
-
2100.01 mg
-
6200.08 mg
51.0001 g -
51000.1 mg
-
220000.8 mg
51.0001 g -
51000.1 mg
-
120000.8 mg
6.20008 g
220.0008 g
120.0008 g

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte

Table of Contents