A&D BA-T Series Instruction Manual page 109

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Outputting external device clock data (clock to be used: external device
clock)
By setting an external device clock for the clock to be used for outputting GLP/GMP data, you can use
the clock data from an external device such as a personal computer or printer for the balance instead of
the internal clock data. This setting is used to unify the clock data with the clock function of the external
device.
Caution
The clock data output from an external device can be used with a device that has a clock function and
that can output the date and time in response to <ESC>D, <ESC>T. (AD-8127 multi-functional compact
printer, RsCom [WinCT] data communication software, etc.)
Output examples of calibration test with the internal weight
The GLP output when the weighing accuracy of the balance is checked using the internal weight is
shown below. (Note that sensitivity adjusment is not performed.)
Clock to be used : Set to [External device clock].
Printer format (AD-8127)
: Space, ASCII 20h
<TERM> : Terminator, C
C
: Carriage return, ASCII 0Dh
R
: Line feed, ASCII 0Ah
L
F
Manufacturer
Model name
Serial number
Calibration test
Zero point result
Loaded weight result
Target weight used
Remarks
Signature
L
or C
R
F
R
Outputs clock data from an external device clock.
ID
Date
Time
109
PC format (RsCom)

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte

Table of Contents