A&D BA-T Series Instruction Manual page 63

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Internal sensitivity adjustment
Display settings 1: MENU key
Int. sensitivity adjustment button
Display settings 2: HOME key
[Internal sensitivity adjustment] screen
1
Back button
Name
Back button
1
Date
2
Time
3
Weight used
4
PRINT button
5
The balance performs sensitivity adjustment using the internal weight.
Be sure to warm up the balance with nothing on the weighing pan for at least an hour, or at least four
hours for BA-6TE/6DTE, with the AC adapter connected to the power supply.
Do not apply vibration and the like to the balance during sensitivity adjustment.
When the adjustment is complete, the [Sensitivity adjustment result] screen for the internal sensitivity
adjustment is automatically displayed.
About the internal weight
The value of the internal weight may change due to factors such as the operating environment and
aging.
Correct the internal weight value as necessary by referring to "Correcting the internal weight value".
In order to maintain the weighing accuracy, it is advisable to perform the external sensitivity adjustment
regularly.
Sensitivity adjustment button
CAL button
Returns to the previous screen.
Displays the date when the internal sensitivity adjustment was complete.
Displays the time when the internal sensitivity adjustment was complete.
Displays the weight used.
Outputs the sensitivity adjustment result.
The result is automatically output if [GLP output] in [Data output] is set to ON.
[Internal sensitivity adjustment] screen
Int. sensitivity adjustment button
Description
63
2
Date
3
Time
4
Weight used
5
PRINT button