A&D BA-T Series Instruction Manual page 77

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

4. Weighing pan check
Make sure that there is nothing on the weighing pan.
1
Skip button
2
Back button
Name
Skip button
1
Back button
2
Confirm button
3
5. Calibration Test (Before Sensitivity Adjustment)
The calibration test is performed before sensitivity adjustment.
When the test is completed, the [Calibration test result] screen is displayed.
When the internal weight is used, this is automatically performed. When an external weight is used,
follow the instruction.
Do not apply vibration and the like to the balance during the calibration test.
1
2
Name
Weight to be
1
used
Back button
2
Result display
3
Skip button
4
Execute button
5
Skips this check item and proceeds to the next item.
Returns to the previous screen.
Considers this check item as OK and proceeds to the next item.
Setting value
(setting range)
Internal weight ,
Select the weight to be used.
External weight
-
Cancels and returns to the previous screen.
Displays the calibration test result.
Skips this check item and proceeds to the next item.
Executes the calibration test.
Description
3
4
5
6
Description
77
3
Confirm button

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte

Table of Contents