A&D BA-T Series Instruction Manual page 121

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Log-in/log-out history
Display settings: MENU key
Name
1
Time
2
User name
3
Level
4
Detail
Back button
5
6
USB output
The history function saves data only in English.
Up to the latest 100 history records are displayed.
Up to the latest 1000 history records are stored and can be output to the USB flash drive as a CSV file.
If the history exceeds 1000 records, the oldest record is deleted and the latest record is added.
Information button
Log-in/log-out history button
1
5
Displays the time when log-in/log-out operations were detected.
Displays the user that performed the detected log-in/log-out operation.
Displays the user level of the user that performed the log-in/log-out operation.
0: Operator
1: Supervisor
Displays "Log-in" or "Log-out".
Returns to the previous screen.
Outputs the history to the USB flash drive as a CSV file.
History button
[Log-in/log-out history] screen
2
3
4
Description
2: Lab manager
121
6
3: Administrator

Advertisement