A&D BA-T Series Instruction Manual page 153

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Command usage examples
In the following command examples, [AK (acknowledge), error code] is set to ON in the [Command
settings] screen described in
(acknowledgment, ASCII 06h) when it processes the command successfully.
ASCII code symbols
C
: Carriage return,
R
: Space,
Example of the ON command (Turning the display on)
Personal computer side
ON command
Example of the R command (Re-zero)
Personal computer side
R command
Example of the CAL command (Sensitivity adjustment with the internal
weight)
Personal computer side
CAL command
"13.11. Command
ASCII 0Dh
ASCII 20h
Time
O N C
L
R
F
Time
R C
L
R
F
Time
C A L C
L
R
F
settings" so that the balance outputs an AK code
L
: Line feed,
F
AK
: Acknowledgment,
AK C
L
R
F
AK C
L
R
F
0.000012
g
AK C
L
R
F
- - - - - - - - -
AK C
L
R
F
0.000012
g
AK C
L
R
F
AK C
L
R
F
153
ASCII 0Ah
ASCII 06h
Balance side
Standby mode
Command received
Process completed
Zero display
Balance side
Before execution
Command received
Waiting for re-zero stability
(processing)
Process completed
Zero display
Balance side
Before execution
Command received
Executing (processing)
Process completed
Result display

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte