8. Repeatability
Load and unload the weight multiple times to obtain the standard deviation.
Follow the instruction to load and unload the weight as many times as the specified measurement count.
The setting in the [Repeatability] screen is common to the standard value setting for the daily/periodic check
settings.
When the measurement is complete, the [Repeatability measurement result] screen is automatically
displayed.
1
2
3
Name
Settings display screen
1
Measurement count
2
Tolerance (standard
3
deviation)
Back button
4
RE-ZERO button
5
Settings button
6
Measure button
7
Skip button
8
Repeatability
measurement result
9
display
Next button
10
4
5
Displays the set measurement count and tolerance (standard deviation).
Set the measurement count.
This is common to
Set the tolerance (standard deviation).
This is common to
Cancels and returns to the previous screen.
Sets the displayed value to zero.
Displays the [Standard value setting] screen. Refer to
setting".
Set the measurement count and tolerance (standard deviation).
Executes the measurement.
Skips this check item and proceeds to the next item.
Displays the repeatability measurement result.
Proceeds to the next item.
6
7
8
9
Description
[11.8.1. Standard value setting:
[11.8.1. Standard value setting:
80
10
Repeatability].
Repeatability].
"11.8. Standard value