External calibration test
Display settings: MENU key
1
2
Name
External weight value input
1
Back button
2
Calibration test result display
3
Confirm button
4
PRINT button
5
Exit button
6
Settings in the red box are default values (factory settings).
*1
Input range
BA-6TE/BA-6DTE
BA-225TE/BA-225DTE
BA-125DTE
Your external weight is used to perform the calibration test.
Be sure to warm up the balance with nothing on the weighing pan for at least an hour, or at least
four hours for BA-6TE/BA-6DTE, with the AC adapter connected to the power supply.
Do not apply vibration and the like to the balance during the calibration test.
When the test is completed, the [Calibration test result] screen is displayed.
Caution
No sensitivity adjustment is performed.
Calibration test/check button
External calibration test button
Input the external weight value. Input range: Input range
This is common to the external weight value input field in the
[Sensitivity adjustment setting] screen.
Returns to the previous screen.
Displays the measurement result.
Confirms the current input setting and proceeds to the next
instruction.
Outputs the measurement result to the device connected to the
balance.
The result is automatically output if [GLP output] in [Data output]
is set to ON.
Exits the screen that shows the measurement result.
: 0.9 g ~
: 9.9 g ~ 200 g ~ 200.1 g
: 9.9 g ~ 100 g ~ 100.1 g
[External calibration test] screen
3
4
5
6
5 g ~
5.1 g