A&D BA-T Series Instruction Manual page 15

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Be sure to warm up the balance before use for at least an hour, or at least four hours for BA-6TE/BA-
6DTE, with the AC adapter connected to the power supply.
Adjust the sensitivity of the balance before using it for the first time or after having moved it to another
location so that accurate weighing can be performed. For how to adjust sensitivity, refer to
"10. [Sensitivity Adjustment]
Do not install the balance where flammable or corrosive gas is present.
How to adjust the level of the balance
Bubble
spirit level
The balance is
The balance is
level (Correct
tilted (Inaccurate
weighing)
weighing)
Level the balance by adjusting the leveling feet
so that the bubble of the bubble spirit level is
centered in the red circle.
When the bubble is off to the left.
Turn the leveling foot on the front right in the
clockwise direction.
When the bubble is off to the
backward position.
Turn both leveling feet on the front in the clockwise
direction at the same time.
Screen".
Up
Leveling foot
Down
Red circle
15
Bubble
Leveling foot
Bubble spirit level
When the bubble is off to the right.
Turn the leveling foot on the front left in the
clockwise direction.
When the bubble is off to the
forward position.
Turn both leveling feet on the front in the
counterclockwise direction at the same time.
Down
Leveling foot
Up