A&D BA-T Series Instruction Manual page 66

Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Automatic sensitivity adjustment
Display settings: MENU key
Automatic sensitivity adjustment button
screen
1
2
Name
Automatic sensitivity
1
adjustment execution
condition setting field
Automatic sensitivity
2
adjustment time setting
Automatic sensitivity
3
adjustment interval
setting
Settings in the red box are default values (factory settings).
This function automatically adjusts the sensitivity of the balance according to ambient temperature
change, set time or interval time using the internal weight. It works even when the display is off. If the
GLP output is enabled, the sensitivity adjustment execution record is output after the sensitivity
adjustment.
The execution condition for the automatic sensitivity adjustment can be based on the temperature
change, preset time or interval.
For preset time, you can configure two settings: "Preset time 1" and "Preset time 2".
For the interval, you can set a value between 0.5h and 24.0h.
Sensitivity adjustment button
Setting value (setting range)
OFF, Temperature change ,
Preset time 1,
Preset time 1&2,
Interval
0:00 - 24:00
0.5 , 01.0, 01.5, 02.0, 02.5,
03.0, 03.5, 04.0, 04.5, 05.0,
05.5, 06.0, 07.0, 08.0, 09.0,
10.0, 11.0, 12.0, 14.0, 16.0,
18.0, 20.0, 22.0, 24.0
[Automatic sensitivity adjustment] settings
3
Set the automatic sensitivity adjustment
execution condition.
If you select the preset time or interval, the
relevant time setting is displayed below the
setting field.
Set the automatic sensitivity adjustment
start time.
Set the automatic sensitivity adjustment
interval.