Minimum weight setting
Display settings: HOME key
Application selection button
HOME key
Minimum weight setting button
1
2
3
1
Minimum weight display
Input directly button
2
Input by measuring repeatability
3
button
Comparison near zero
4
Outputting data less than the
5
minimum weight
Output the minimum weight button
6
Settings in the red box are default values (factory settings).
*
Near zero is within ± 12 d of 0 mg. "d" is a unit of readability.
This is the screen for setting the minimum weight.
Minimum weight input
Display settings: HOME key
Application selection button
HOME key
Minimum weight setting button
Input directly button
1
Minimum weight input
APP button
Weighing screen with the minimum weight alert function
Name
APP button
Weighing screen with the minimum weight alert function
Select [Minimum weight alert function].
[Minimum weight setting] screen
Setting
value
-
Displays the currently stored minimum weight.
Displays the minimum weight input screen.
Displays the [Minimum weight setting] screen for
measurement.
Select whether to include or exclude near zero for
Include
Exclude
the minimum weight alert
If set to OFF, data with the weighing value less
OFF
ON
than the minimum weight is not output.
Outputs the minimum weight.
Select [Minimum weight alert function].