Capillary Cleaning And/Or Replacement; Figure 7-25. Remove Capillary, Disconnect And Unfasten - Thermo Scientific 42iQ Instruction Manual

No-no2-nox analyzer

Advertisement

Capillary
Cleaning and/or
Replacement
Thermo Scientific
Use the following procedure to clean or replace the capillary.
Equipment required:
Phillips drive, #2
Hex drive, 7/64
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2–1).
2. Disconnect the plumbing.
3. Using #2 Phillips drive, unfasten captive hardware.
Figure 7–25. Remove Capillary, Disconnect and Unfasten
4. Slide capillary plate upwards clearing the partition panel keyway.
Servicing

Capillary Cleaning and/or Replacement

42iQ Instruction Manual
7-23