Oxygen Sensor Capillary Replacement; Figure 7-64. Replacing The Oxygen Sensor Capillary - Thermo Scientific 42iQHL Instruction Manual

High level no-no2-nox analyzer

Advertisement

Oxygen Sensor
Replacement
Thermo Scientific
Use the following to replace the oxygen sensor capillary.
Capillary
Equipment required:
Phillips drive, #2
1. Remove the oxygen sensor DMC from the instrument. Refer to
"Oxygen Sensor
2. Unlock tube clamp.
3. Unplug tube after unlocking tube clamp.
4. Using a #2 Phillips drive, unfasten four #6-32 flat head screws.
Figure 7–64. Replacing the Oxygen Sensor Capillary
5. Pull capillary assembly outwards and apart as shown.
Removal" on page 7-56.
Servicing
Optional DMC Oxygen Sensor
42iQHL Instruction Manual
7-61