Optical Bench Cleaning; Figure 5-3. Absorption Cleaning Removing Nuts; Figure 5-4. Absorption Tube Removal - Thermo Scientific 49iQ Instruction Manual

Ozone analyzer

Advertisement

Maintenance

Optical Bench Cleaning

Optical Bench
Cleaning
5-6
49iQ Instruction Manual
Use the following procedure to clean the mirrors.
Equipment required:
Hex drive, 7/64
1. Unfasten nuts with hands. No tool required.
Figure 5–3. Absorption Cleaning Removing Nuts
2. Pull up then pull out the absorption tube.
Figure 5–4. Absorption Tube Removal