DMC Optical
Bench Removal
Optical Bench
Assembly Removal
43iQTL Instruction Manual
Use the following to remove the DMC optical bench from the instrument.

Optical Bench Assembly Removal

G
PMT Replacement
G
Flasher Pack Replacement
G
Flasher Pack Lamp Replacement
G
Photo Lamp Detector Board Replacement
G
Kicker Assembly Replacement
G
Use the following procedure to remove the optical bench assembly.
Equipment required:
Phillips drive, #2
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2–1).
2. Unplug two DMC cables.
3. Disconnect all plumbing.
4. Unfasten six captive hardware.