Editing the Current Method
TIMED EVENTS SCREEN
To move to the TE1 data field, press the
4
C, or M. These letters indicate the following:
O: Open - The relay opens at the time of the event.
C: Close - The relay closes at the time of the event.
M: Momentary - The relay closes for about two seconds, then opens. If the previous
state of the relay is Close, this action may not cause the desired operation. Momentary
closures should be preceded with an Open event.
To add another event, press either the
5
will be added and the Time field for the next event will increment by 0.01 minute. You
can add up to 20 events using this process.
To save your entries, press the
6
Screen and saves the changes until you turn the power off on the Detector. When you
restore power, the default values for the method will also be restored. To permanently
store the changes, refer to the "Storing a Method" section in this chapter.
key. The timed event may be either O,
enter
key or the
enter
key. This also takes you back to the Method
return
Using Method Mode
key. A second event
insert
4 - 11