Series 200 Operating Precautions - PerkinElmer 200 UV/VIS Series User Manual

Table of Contents

Advertisement

Safety and Regulatory Information

Series 200 Operating Precautions

To protect yourself from harm and to prevent malfunctioning of the system, please observe
the following guidelines when operating the UV/VIS Detector:
• Before using the instrument, pay attention to the warnings presented in the beginning of
the manual.
• Wear the appropriate protective clothing, including safety glasses and gloves, when pre-
paring samples and solutions for use with this instrument.
• Always follow the correct safety procedure and the manufacturer's recommendations
when using solvents.
• Keep solvents with a boiling point less than 30 °C under helium (10 kPa) in an optional
solvent cabinet (for example, the SEC 4).
• Do not keep buffer solutions in the system when the instrument is not in use, since salt
may crystallize out.
• Do not use acidic chlorides with pH less than 7; they can cause corrosion. Chloroform and
tetrachloromethane decompose slowly and form HCl.
• Do not use chloride TRIZMA buffers since they form HCl in equilibrium. Use phosphate
buffers instead.
• Do not use inorganic acids, concentrated organic acids or acidic buffer solutions.
• Ensure that the system is flushed with degassed solvent.
• Ensure that the other instruments in the system are ready to run.
• Ensure that all connecting communications and power cables are correctly installed for
your analysis.
• It is possible to clean the exterior surfaces of the instrument with a soft cloth dampened
with a mild detergent and water solution. Do not use abrasive cleaners or solvents.
Safety and Regulatory Information
S - 15