Setting Configuration Options
• LEAK
There are multiple Leak error messages:
INT LEAK SENSR SHRTD (Internal Leak Sensor Short)
INT LEAK SENSR OPEN (Internal Leak Sensor Open)
INT LEAK (Internal Leak)
EXT LEAK SENSR SHRTD (External Leak Sensor Short)
EXT LEAK SENSR OPEN (External Leak Sensor Open)
EXT LEAK (External Leak)
A Short or Open error on the Internal Leak Detector implies an error in the leak detector
electronics and you should contact your local PerkinElmer service representative.
A Short or Open error on the External Leak Detector may be caused by incorrect
installation of the leak detector. Check the installation. If all appears well
then the error is probably in the leak detector electronics in the detector and you should
contact your local PerkinElmer service representative.
An Internal or External Leak is caused by solvent contacting the leak detector. Power
down your complete system, determine the location and cause of the leak and correct
before restarting your system.
Follow the procedure below to check the instrument log:
From the Main Screen, press F7
1
MAIN SCREEN
CNFG
CONFIGURE SCREEN
to move to the Configure Screen.
F4
Configuring the Detector
F7
6 - 19