Editing Timed Events - PerkinElmer 200 UV/VIS Series User Manual

Table of Contents

Advertisement

Using Method Mode

Editing Timed Events

Timed events control the relay contact closures which in turn control auxiliary instruments
(fraction collector, autosampler, etc.). You can actuate two timed events (TE1 and TE2) a
maximum of twenty times while running the method.
To edit the timed events, follow the procedure below.
From the Method Screen, press F2
1
METHOD EDIT SCREEN
Press F5
2
highlighted.
Using the numeric keys on the keypad, type in the time at which you want the first
3
timed event to occur.
NOTE: If you make a mistake entering numeric values in any of the fields on this
screen, press the
4 - 10
to move to the Timed Events Screen. The data field for the Time is
T.E.
TIMED EVENTS SCREEN
key and retype the number.
clear
to display the Method Edit Screen.
EDIT
F5
Editing the Current Method