Setting Operating Parameters for Immediate Change
Press the right arrow key or the
4
on the detector's configuration, either RSP or PW is displayed for this field. You use
this field to select a value for the response time or peak width at half height. To change
the value, press the corresponding softkey. For more information on which value to
select, see the "Detector Response Time" topic in Chapter 8, Principles.
Press the right arrow key or the
The relay event may be either O, C, or M. These letters indicate the following:
5
O: Open - The relay opens immediately.
C: Close - The relay closes immediately.
M: Momentary - The relay closes for about two seconds, then opens.
If the previous state of the relay is Close, this action may not cause the desired opera-
tion. Momentary closures should be preceded with an Open event.
Press the cursor arrow key or the
6
Select a value for TE2. The relay event may be either O, C, or M, as explained in Step
5.
key to move to the RSP/PW field. Depending
enter
key to set the RSP and move to the TE1 field.
enter
key to set TE1 and move to the TE2 field.
enter
Using Manual Mode
3 - 7