About Sequences
The following table describes the parts of the Sequence Screen.
Part of the
Number
Screen
1
SET
2
METH
3
Sequence Status
4
INJ
The Sequence Screen displays the following softkeys:
Softkey
LINK
Pressing this softkey selects the sequence for operation. When the
Run key is pressed, the sequence will be run.
DELS
Pressing this softkey deletes the entire sequence.
Description
Indicates a line in the sequence. There can be from
one to 20 lines in the sequence.
The method number for the Set. The method must
have one or more steps.
"LINKD" indicates that the sequence has been set
up and is ready to run. In the figure above, the
sequence status is S03.001. The "S" indicates
"Sequence." The "03" indicates the current set, and
"001" indicates the current injection.
The number of injections the autosampler will
make from each vial in the set. The range is from
zero to ninety-nine.