Configuring the Detector
Setting Configuration Options
Checking the Wavelength Calibration
This section describes how to run the wavelength calibration check. You can run the wave-
length calibration check when you suspect that you are getting incorrect results from the
λ
detector. When you select
CHK from the Configure menu, the Detector displays the Wave-
length Calibration Check Screen. This screen allows you to run a check on the wavelength
calibration. Before starting the check, you must replace the flowcell with a Holmium Oxide
Filter. For information on replacing the flowcell, see Chapter 7, Maintenance.
MAIN SCREEN
F7
From the Main Screen, press F7
to move to the Configure Screen.
CNFG
1
CONFIGURE SCREEN
F3
6 - 16