Checking The Wavelength Calibration - PerkinElmer 200 UV/VIS Series User Manual

Table of Contents

Advertisement

Configuring the Detector
Setting Configuration Options

Checking the Wavelength Calibration

This section describes how to run the wavelength calibration check. You can run the wave-
length calibration check when you suspect that you are getting incorrect results from the
λ
detector. When you select
CHK from the Configure menu, the Detector displays the Wave-
length Calibration Check Screen. This screen allows you to run a check on the wavelength
calibration. Before starting the check, you must replace the flowcell with a Holmium Oxide
Filter. For information on replacing the flowcell, see Chapter 7, Maintenance.
MAIN SCREEN
F7
From the Main Screen, press F7
to move to the Configure Screen.
CNFG
1
CONFIGURE SCREEN
F3
6 - 16