Maintenance
Avertissement
5. Unscrew the knurled screw that is holding the front panel in place.
6. Swing the front panel open to the right.
7. Slide the drawer out to the first stop.
8. Loosen the Allen bolt on the lower left side of the drawer with the special Allen wrench
provided. (See picture below.)
9. Slide the drawer completely out of the cover.
10. Unfasten the lamp retainer spring.
11. Grip the entire lamp with one hand while holding the detector steady with the other.
Slide the entire lamp assembly off the lamp mount.
12. Unplug the lamp.
Allen bolt
Figure 7-5.
Preparing to slide the drawer out of the cover
7 - 12
WARNING Hot
Replace lamp with PerkinElmer part
number N292-0149.
LA LAMPE EST TRÈS CHAUDE!!
Remplacer la lampe par une lampe
PerkinElmer, numéro de pièce N292-0149.
Do not disturb the black fan connector
located below the white lamp connector. If
the fan connector is not properly connected,
the detector will not operate.