Setting Configuration Options
Press F3
2
instructs you to remove the flowcell from the Detector and replace it with a Holmium
Oxide Filter. For information on the procedure for removing and replacing the flow-
cell, see Chapter 7, Maintenance.
Press the
3
checks for the Holmium Oxide Filter and then runs the calibration test.
When the test is complete, the Detector displays the results of the test and instructs
4
you to remove the Holmium Oxide Filter. Press the
figure Screen.
If the wavelength calibration consistently produces incorrect values, contact your
PerkinElmer service representative. The detector can only be recalibrated by a
PerkinElmer engineer. Your service representative can perform monthly checks on
the wavelength calibration if your lab requires it.
λ
to move to the Wavelength Calibration Check Screen. The screen
CHK
WAVELENGTH CALIBRATION CHECK SCREEN
softkey to run the wavelength calibration check. The Detector
STRT
WAVELENGTH CALIBRATION - RESULTS SCREEN
Configuring the Detector
key to return to the Con-
return
6 - 17