2.3 Vertical and Horizontal Angle Tilt Correction

When the tilt sensors are activated, automatic correction of vertical and horizontal angle for
mislevelment is displayed.
To ensure a precise angle measurement, tilt sensors must be turned on. The display can also be used
to fine level the instrument. If the (TILT OVER) display appears the instrument is out of automatic
compensation range and must be leveled manually.
Zenith
Standing axis
ES compensates both the vertical angle and the horizontal angle readings due to inclination of the
standing axis in the X and Y directions.
V :
°
'
HR:
°
'
<X TILT OVER>
Standing Axis in the X direction
out of range
The display of Vertical or Horizontal angle is unstable when instrument is on an unstable stage or a
windy day. You can turn off the auto tilt correction function of V/H angle in this case.
To set auto tilt correction from the moment that power is on, See Section 6.4.3 "Vertical and
Horizontal Angle Tilt correction (Tilt ON/OFF)".
Inclination of the standing
axis in the X direction
When the instrument is out of compensation. (TILT OVER)
V :
"
HR:
"
<Y TILT OVER>
Standing Axis in the Y direction
2 PREPARATION FOR MEASUREMENT
Standing axis
Horizontal
°
'
"
°
'
"
out of range
2-3
Zenith
Inclination of the standing
axis in the Y direction
Trunnion axis
V :
°
'
"
HR:
°
'
"
<XY TILT OVER>
Standing Axis in the X and Y
directions out of range