ERROR DISPLAYS

The following is a list of the error messages displayed by the instrument and the meaning of each
message.
There is a possibility that an unexpected error occasionally occurs when using an old battery. In such a
case, exchange it for a new battery, and please check the operation of the instrument.
If error still persist after attempting to clear them, contact your local dealer.
Error code
When area calculating, there are less than 3
3 points
required
points coordinate data in selected file.
The air is shimmering a lot, etc., measuring
conditions are poor.
The center of the target cannot be sighted.
BadCondition
Unsuitable distance measurement conditions
when reflectorless measurement is set.
When reflectorless measurement is set,
distance cannot be measured because the
laser beam is striking at least two surfaces at
the same time.
CALC
Calculation is impossible from the data input.
ERROR
When deleting coordinate data, it can not
DELETE
ERROR
be done successful.
Measuring conditions are poor, and there is
SIGNAL
not enough reflective light for measuring
OFF
distances.
E020's,
EDM error
E031, E033
Displayed when REM measurement carried
E35
out to the range from zenith or nadir ±6°.
Displayed when Vertical angle -position is
E72
adjusted in wrong position.
The instrument was not leveled when Vertical
E73
angle 0-position is adjusted.
Mainly at the time data transmission between
E80's
ES and external instrument.
E097
EDM device error
E123, E124
Tilt error
E198
Angle measurement error
EDM distance measurement time out error
E201
Communication with EDM is poor
E205
FILE
The same file name exists.
EXISTS
Description
24-1
24 ERROR DISPLAYS
Countermeasures
Confirm the file data and calculate again.
Perform measurement again when
measuring condition becomes better.
Resight the target.
Choose a single surface target for distance
measurement.
Confirm the input data.
Confirm the data and delete again.
Resight the target. When using reflective
prisms, effectiveness will be improved by
increasing the number of prisms used.
Repair is required.
Operate in the range out of ± 6° from the
zenith or nadir.
Repair is required.
Level the instrument then carry the
adjustment work.
Confirm operation procedure is correct or the
connection cables are correct.
Repair is required.
Contact your local dealer.
Contact your local dealer.
If the same error message is repeated, repair
is required.
If the same error message is repeated, repair
is required.
Use another file name.