Installation Of The Case Assembly (E7044Kf) For Calibration Gas Cylinder - YOKOGAWA ZR402G User Manual

Integrated type zirconia oxygen/humidity
Hide thumbs Also See for ZR402G:
Table of Contents

Advertisement

Horizontal mounting on the ZR202G
Vertical mounting on the ZR202G
Figure 3.10
3.4
Installation of the Case Assembly
(E7044KF) for Calibration Gas Cylinder
The case assembly is used to store the G7001ZC zero gas cylinders.
The following should be taken into consideration:
(1) Easy access for cylinder replacement
(2) Easy access for checking
(3) Near to the detector and converter as well as the flow setting unit.
(4) The temperature of the case should not exceed 40°C due to rays of the sun or radiated
heat.
(5) No vibration
(-A)
214
44
MAX
Span gas inlet
Rc1/4 or 1/4NPT(Female)
(-B)
180
44
MAX
Automatic Calibration Unit Mounting
Display side
40
40
66.5
166.5
Span gas inlet
Rc1/4 or 1/4NPT(Female)
Reference gas inlet
Rc1/4 or 1/4NPT(Female)
40
<3. Installation>
Unit: mm
Terminal box side
Zero gas inlet
Rc1/4 or 1/4NPT(Female)
Reference gas inlet
Rc1/4 or 1/4NPT(Female)
166.5
Zero gas inlet
Rc1/4 or 1/4NPT(Female)
40
66.5
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
3-7
F3.15E.ai