Software Configuration Checks; System Setup; Measure Setup - GE Voluson i BT06 Service Manual

Table of Contents

Advertisement

GE
D
KTI106052, R
IRECTION
Section 4-6

Software Configuration Checks

4-6-1

System Setup

Press the UTILITIES key and select the SYSTEM item from the "Utilities" menu - which is shown in the
menu area on the left side of the screen. The System Setup desktop offers different pages to check:
Table 4-10
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
4-6-2

Measure Setup

Press the UTILITIES key and select the MEASURE item from the "Utilities" menu - which is shown in
the menu area on the left side of the screen.
The Measure Setup desktop offers three different pages to check.
Parameters and possible adjustments depend on the selected Application.
NOTICE
To view, add, delete, reorder, edit or when creating a new parameter (in the
!! NOTICE:
is very important that all items are chosen correctly and that the relevant item is highlighted.
For further information refer to Chapter 17 in the Basic User Manual of Voluson i / Voluson e.
Table 4-11
Step
1
2
3
4-38
10
EVISION
Software Configuration (System Setup) Checks
Task
General: Check Date and Time setting
General: Check that Location (Clinic Name) is correct
General: Check Language settings
General: Check EUM Language settings
General: Check "Display" settings
User Settings: Adjust/check "LCD Brightness"
User Settings: Check settings of the "Patient ID" fields
User Setting: Check all the User Settings
P1-P2-P3: Check assignment of the [Print] tab
P1-P2-P3: Check assignment of the [Save] tab
P1-P2-P3: Check assignment of the [Export] tab
Option: Check that all options are set up correct
Network:
®
Check DICOM
, Sonoview and Network configuration
Measurement Setup Checks
Task
Measure & Calc:
Check all settings for all applications
Application Parameters:
Check all settings for all applications
Global Parameters: Check all settings
Section 4-6 - Software Configuration Checks
Expected Result(s)
Date and Time are correct
Location Name is correct
desired Language is displayed
desired EUM Language is displayed
"Display" settings assigned as desired by the customer
setup displays a complete gray scale
settings assigned as desired by the customer
settings assigned as desired by the customer
settings assigned as desired by the customer
settings assigned as desired by the customer
settings assigned as desired by the customer
D = Demo , I = Inactive , P = Permanent
settings assigned as desired by the customer
MEASURE & CALC
Expected Result(s)
setting assigned as desired by the customer
setting assigned as desired by the customer
setting assigned as desired by the customer
V
i / V
e
OLUSON
OLUSON
S
M
ERVICE
ANUAL
page), it