Connecting the DVD+R/RW Drive

3-5-7-1

Connection directly to the Voluson i / Voluson e system

The DVD+R/RW Drive can be connected to the USB ports on the rear connector panel.
isolation transformer
When connecting directly to the system, it is necessary to use additional AC/DC power supply (supplied
CAUTION
with the drive). As a safety requirement, the AC/DC must be medical grade (conforming 60601-1
!! CAUTION:
standard) or the safety must be ensured by the use of a medical grade isolation transformer.
Auxiliary equipment with direct mains connection requires galvanic separation e.g. by use of a medical
NOTICE
grade isolation transformer.
!! NOTICE:
NOTE:
The connection Scheme for DVD Drive Samsung, De-Lock and Teac is similary the same.
3-46
10
EVISION
NOTICE: Auxiliary equipment with direct mains
connection requires galvanic
separation e.g. by use of a medical
grade isolation transformer.
Power Supply
power
socket
Figure 3-60 Connection Scheme - DVD+R/RW Drive