GE Voluson i BT06 Service Manual page 140

Table of Contents

Advertisement

GE
D
KTI106052, R
IRECTION
3-6-2
Power On / Boot Up
NOTE:
Through the protection plate near the plug, the power cable can only be plugged in one way.
L-shaped circular plug
of main power cable
connected to DC IN
4.) If not already done, connect peripheral devices such as DVD Drive, etc.
Links to connection schemes can be found in
5.) Connect the other end of the main power cable to a hospital-grade main power outlet with the
proper rated voltage.
When AC power is applied to the scanner, the On/Off button on the control console illuminates amber,
NOTICE
indicating that the System (including the Back-end Processor) is in Standby mode.
!! NOTICE:
6.) Hold down the On/Off button (see:
The system automatically performs an initialization sequence which includes the following:
-
-
-
As soon as the software has been loaded, the system enters 2D-Mode with the probe and application
that were used before the system was shut down.
NOTE:
Total time used for start-up is typically one minute.
NOTE:
When turning on the Voluson i / Voluson e from Standby mode, it takes a few seconds before it
responds.
Do not push the ON/OFF button again during this period. A second push will initiate a full shutdown.
3-70
10
EVISION
(cont'd)
Figure 3-86 main power cable connected to Voluson i / Voluson e
Loading the operating system.
Running a quick diagnostic check of the system.
Detecting connected probe.
Figure 3-87 On/Off Button on Control Panel
Section 3-6 - Completing the Set Up
Table 3-9, "Connection Procedures," on page
Figure
3-87) on the control panel for ~3 seconds.
V
i / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
protection plate
3-23.
On/Off Button
e