Software Configuration Checks - GE LOGIQ 5 Service Manual

Hide thumbs Also See for LOGIQ 5:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300000, R
IRECTION
EVISION
4-3-12
ECG Check Out (this option is currently not available on the LOGIQ™ 5
Connect the ECG Harness and check:
Table 4-5
Step
1.
Section 4-4

Software Configuration Checks

Table 4-6
Step
1.
2.
3.
4.
5.
2
ECG Harness Check
Task to do
Connect the ECG at the Connector on the Front of the
scanner.
Software Configuration Checks
Task to do
Check Date and Time setting
Check that Location (Hospital Name) is correct
Check Language settings
Check assignment of Printer Keys
Check that all of the customer's options are set up correct All authorized functions are enabled
Chapter 4 Functional Checks
LOGIQ™5 S
Expected Result(s)
It will display a curve along the bottom edge of the image
sector
Expected Result(s)
Date and Time are correct
Location Name is correct
Desired Language is displayed
Print1-4 Keys are assigned as desired by the customer