GE
D
KTI106052, R
IRECTION
8-18-3
Replacement of the Fuses
8-18-3-1
Manpower
One person, 10 minutes
8-18-3-2
Tools
Phillips screwdriver 1 and 2
8-18-3-3
Preparations
1.) Power Off/Shutdown the system as described in
8-18-3-4
Fuses - Removal Procedure
1.) Open the fuse protection on the rear of the isolation transformer.
2.) Pull out the fuse holder and remove the 2 fuses.
Fuse holder
(e.g. 230V)
8-18-3-5
Fuses - Installation Procedure
1.) Insert 2 new Fuses.
2.) Push the fuse holder back into position (correct voltage should be shown in upper line
e.g, 230V, see:
3.) Reconnect the mains power cable.
8-42
10
EVISION
Figure 8-57 open fuse protection on rear of transformer
Figure 8-58 change fuses of isolation transformer
Figure
8-58, above) and then close protection.
Section 8-18 - Replacement of the Modo Cart Components
Section 3-6-3 on page 3-72
Fuse holder
(back)
V
OLUSON
S
ERVICE
Fuse
Fuse
i / V
e
OLUSON
M
ANUAL