GE
D
KTI106052, R
IRECTION
3-5-3
Connection and Usage of the optional Voluson Station
DO NOT CONNECT any devices to any AC power other than the on-board outlets from
WARNING
WARNING
the transformer, as this might increase leakage currents and sacrifice patient safety.
!! WARNING:
WARNING
WARNING
Before starting the Voluson i / Voluson e or Peripherals that are connected to the
Voluson Station, ALWAYS make sure that rating of the power supply fuses is
!! WARNING:
compatible and the voltage output selector is in the correct position.
The wrong fuses and position of the voltage output selector may cause major damage
on the Voluson i / Voluson e system and connected peripherals.
Make sure that all peripherals are firmly secured in position according installation procedure.
CAUTION
!! CAUTION:
3-5-3-1
Attaching the Voluson i / Voluson e to the Voluson Station
A battery or an empty battery shell should always be attached to the Voluson i / Voluson e system
NOTICE
before attaching it to the Voluson Station.
!! NOTICE:
1.) Place the system in the docking bay on the top shelf of the Voluson Station (1)
2.) Pull down the docking lever(2+3).
docking lever
10
EVISION
Figure 3-38 Voluson i / Voluson e attached to the Voluson Station
Chapter 3 - Setup Instructions
V
i / V
e
OLUSON
OLUSON
S
M
ERVICE
ANUAL
.
3-35