Software Configuration Checks; Peripheral Checks - GE LOGIQ e Basic Service Manual

Hide thumbs Also See for LOGIQ e:
Table of Contents

Advertisement

General Procedures and Functional Checks

Software Configuration Checks

Step
1.
Check Date and Time setting
2.
Check that Location (Hospital
Name) is correct
3.
Check Language settings
4.
Check assignment of Printer Keys
5.
Check that all of the customer's
options are set up correct

Peripheral Checks

Color Printers
4-52
Table 4-10:
Software Configuration Checks
Task to do
This section describes the final setup for the Peripherals and the
Operational Check-out. For Peripherals installation information,
See 'Peripherals Installation' on page 3-34 for more information.
Check that peripherals work as described below:
The Printers are controlled from the P2 on the system's control
panel. The factory setting is P2 for the standard print.
Take Sony UP-D25MD Color Printer for an example to show the
control assignment procedure.
Expected Result(s)
Date and Time are correct
Location Name is correct
Desired Language is displayed
For LOGIQ e, the default function for Print1-3
Keys is P1 (store image); P2 (print); P3 (USB
Quick Save). Print1-3 Keys can also be
assigned as desired by the customer
All authorized functions are enabled
LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6

Advertisement

Table of Contents
loading

This manual is also suitable for:

548351854835225498957549958254835205483521