Network Troubleshooting; No Connection To The Network At All; Ge Remote Service Connection - GE Voluson i BT06 Service Manual

Table of Contents

Advertisement

GE
D
KTI106052, R
IRECTION
7-8-6

Network Troubleshooting

7-8-6-1

No Connection to the Network at All

1.) Check that the network cable between the scanner and the wall network is connected and well
seated in both ends. (Use a network cable that is known to be OK.)
2.) Connect a network cable between your Scanner and your PC. Try to ping from the Scanner to the
IP address on the PC. If OK, the hardware connection inside the Scanner is OK.
7-8-6-2

GE remote service connection

The customer gets visual information about the network status (shown in the status area on the right
side of the screen). The network status is defined as follows:
NO ICON - Cable disconnected or no network signal on a connected cable.
GREEN
- All the following processes are active: UL_VNC, UL_Telnet or UL_CSD.
BLUE
These processes are used when remote access is active, so they are used as an indicator.
NOTE:
The status is checked every 10 seconds, regardless whether you're in read or write mode.
7-26
10
EVISION
- Cable connected to a network. Does not imply proper network settings.
Section 7-8 - Troubleshooting Trees and Instructions