GE Voluson i BT06 Service Manual page 208

Table of Contents

Advertisement

GE
D
KTI106052, R
IRECTION
Pre-Volume Mode Functions (cont'd)
4-4-5-1
Table 4-7
Step
1
3D RENDERING
2
SECTIONAL PLANES
3
TUI*
4
4D REAL TIME
RT 4D BIOPSY*
5
BT-Version:
VCI*
6
BT-Version:
STIC*
7
BT-Version:
SONOVCAD HEART*
8
BT-Version:
SONOVCAD LABOR
9
BT-Version:
SONOAVC
10
BT-Version:
11
12
13
14
Volume Box Position
15
and Volume Box Size
16
QUALITY
17
VOL. ANGLE
18
Start Acquisition
4-16
10
EVISION
Pre-Volume Mode Functions
Task
3D Static volume acquisition + rendered 3D image (also in combination Color or PD Mode)
3D Static volume acquisition resp. 4D volume acquisition without rendered 3D image
This method of visualization is consistent with the way other medical systems such as CT or
MRI, present the data to the user (slices through the data set, which are parallel to each other).
Real Time 4D - continuous volume acquisition and parallel calculation of 3D rendered images
Real Time 4D Biopsy
continuous volume acquisition and parallel calculation of 3D rendered images
BT Version:
This feature is only available at Voluson i systems with BT´07 software and higher.
Volume Contrast Imaging - improves the contrast resolution and the signal / noise ratio and
therefore facilitates finding of diffuse lesions in organs
BT Version:
This feature is only available at Voluson i systems with BT´07 software and higher.
The fetal heart or an artery can visualized in 4D (also in combination with PD, HD-Flow or CFM)
BT Version:
This feature is only available at Voluson i systems with BT´09 software.
VCAD is a technology that automatically generates a number of views of the fetal heart to make
diagnosis easier. At this time it can help to find the right and left outflow tract of the heart and
the fetal stomach. By aligning the 4 Chamber view of the heart to a set diagram the system will
BT Version:
generate views of LVOT, RVOT, Stomach, IVC/SVC and Ductal arch
This feature is only available at Voluson i systems with BT´09 software.
This Feature allows for supervision of labor using specific measurements aided by onscreen
orientation marks.
BT Version:
This feature is only available at Voluson i / Voluson e systems with BT´09 software.
This Feature can automatically detect low echogenic objects (e.g., follicles) in a volume of an
organ (e.g., ovary) and analyze their shape and volume
BT Version:
This feature is only available at Voluson i / Voluson e systems with BT´09 software.
CLASSIC: Advanced Mode with dedicated Voluson Volume features
SMART: Easy/Speedy operation with as few controls as possible
- Quarter size display of Sectional Planes without 3D image or
- Quarter size display of Sectional Planes + rendered 3D image
(Note: The display depends on selected Acquisition- and Visualization Mode!)
Dual size display of Sectional Planes + rendered 3D image.
(Note: The display depends on selected Acquisition- and Visualization Mode!
This format is not possible for Static 3D Acquisition)
- Full size display of a the reference image or
- Full size display of the rendered 3D image.
(Note: The display depends on selected Acquisition- and Visualization Mode!)
Adjust the POS (Position) resp. SIZE of the Volume Box (ROI = Region Of Interest) with the
trackball in the 2D Single image.
The upper trackball key to change the trackball function from Box Position to Box Size.
Changes the line density against the acquisition speed (low, mid1, mid2, high1, high2).
To select the Volume Sweep Angle.
Press the FREEZE key resp. the right trackball key to start the Volume acquisition.
Section 4-4 - Functional Checks
V
Expected Results
i / V
e
OLUSON
OLUSON
S
M
ERVICE
ANUAL

Advertisement

Table of Contents
loading