Recalibrating The Battery; Recalibration By Means Of Internal Battery Charger - GE Voluson i BT06 Service Manual

Table of Contents

Advertisement

GE
D
KTI106052, R
IRECTION
Table 3-12 Battery Charger LED Indicators and Status
-
-
Remove the battery from the charger unit battery bay and install it in the Voluson i / Voluson e system.
NOTE:
For more details, refer to
3-6-4-2

Recalibrating the Battery

Recalibration will take between 9 an 14 hours (depending on state of charge of the battery).
NOTICE
In order to carry out a full recalibration the following must occur:
• a fully battery charge (to take it to 100% charged and establish current "Full point")
• a fully calibration discharge (at this point the battery is calibrated, but also empty)
• a fully recharge to return it to use
A calibration cycle will be faster if the battery is fully charged to begin with.
There are two possibilities to recalibrate the Voluson i / Voluson e´s lithium ion battery:

Recalibration by means of Internal Battery Charger

Recalibration by means of External Battery Charger
NOTE:
For more details on recalibration and an explanation of why it is necessary, refer to
Section 5-5-6-2 "Recalibration" on page
3-76
10
EVISION
LED Indicator
Green Flashing
Green Solid
Blue flashing
Blue Solid
Red Flashing
Red Solid
Recharge Time:
Time for a full charge from 0% to 100% state of charge is typically 4.5 hours.
Recalibration Time
minimum recalibration time: approximately 9 hours
maximum recalibration time: approximately 14 hours
see:
Section 3-6-4-2 "Recalibrating the Battery" on page
Section 5-5-6 "Battery Charger" on page
Section 3-6 - Completing the Set Up
Status
Battery charging
Battery fully charged
Battery in calibration mode
Battery fuel gauge calibrated
Battery fuel gauge in need of recalibration
Error
5-34.
V
i / V
OLUSON
S
ERVICE
3-76.
5-33.
e
OLUSON
M
ANUAL