Power Supplies for additional equipment MUST comply with IEC 60601-1. Auxiliary equipment
CAUTION
with direct main connection requires galvanic separation of the signal and/or control leads.
!! CAUTION:
WARNING
WARNING
DO NOT attempt to use different peripherals and accessories other than approved and
provided by GE!
!! WARNING:
Any unauthorized peripherals may cause system failure or damage!
After each installation, the leakage currents have to be measured according to
WARNING
WARNING
IEC 60601-1 respectively UL 60601-1 (see:
!! WARNING:
Please observe that some printers may not be medical devices! If they are not medical devices,
CAUTION
they have to be located outside of the patient environment (acc. to IEC 60601-1 / UL 60601-1).
!! CAUTION:
NOTE:
For more detailed Safety Considerations when connecting auxiliary devices to the Voluson i / Voluson
e system, please review:
3-24
10
EVISION
Chapter 1 - Auxiliary Devices Safety on page
Section 3-5 - Connection of Auxiliary Devices
Section 10-7-2 on page
1-10.