Attaching the Voluson i / Voluson e to the Voluson Station (cont'd)
3-5-3-1
If applicable, connect the Probe Mux Box.
3.) Push the probe connector into the probe socket of the Voluson i / Voluson e.
4.) Lift the system's locking lever upwards to lock the probe connector.
The Probe Mux Box enables the connection of up to 3 probes to the Voluson i / Voluson e.
5.) Connect the cable to the rear of the Voluson i / Voluson e. The cable is listed below:
LAN cable - Connect the ethernet cable, routing from the LAN isolation box, into the network port
of the Voluson i / Voluson e.
6.) Mount and connect other peripherals as described in
7.) Connect the main power cable to the Power Supply of the Voluson Station; the other end connect
to a hospital-grade main power outlet with the proper rated voltage.
3-36
10
EVISION
Figure 3-39 Voluson i / Voluson e attached to the Voluson Station
Section 3-5 - Connection of Auxiliary Devices
V
OLUSON
Section 3-5-3-8
and
Section
i / V
e
OLUSON
S
M
ERVICE
ANUAL
3-5-3-9.